Ecocell 404
convincingly economical laboratory drying oven
chamber volume 404 litres
temperature range from 5°C above ambient to 250/300°C
natural fine airflow in the work area, noiseless operation
high-quality and safe drying, heating and tempering of common laboratory materials in the laboratory and industrial areas
high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
output for connecting a printer or PC
possibility of validation (IQ, OQ)
Technical parameters
Interior of stainless steel material No. 1.4301
volume 404 l
depth 520 mm
width 540 mm
height 1415 mm
Tray
standard equipment 2 pcs
storage area (w × d) 520 × 485 mm
racks 19 max. no
distance between guide rails 70 mm
Maximal weight of the load *)
inside the oven 100 kg/oven
per tray 30 kg/tray
Door
1 pcs
External dimensions (including door and handle)
depth 790 mm
diameter of the air branch outer/inner 52/49 mm
width 760 mm
1910 (K)
Package dimensions (three layers carton)
depth 860 mm
width 830 mm
depth 2085 mm
Weight
brut 175 kg
nett 150 kg
Working temperature (beginning of the regulation)
from 5 °C above ambient temp. to: 250 °C
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door
time variation cca (±) 1 °C
space deviation cca (±) % of the reached temperature 2,5 %
Time required to reach 250 °C with closed air flap and voltage 230 V
85 min
Heat emission at 250 °C
1940 W
Air exchange speed at 250 °C
4 per hour
Electricity - mains 50/60 Hz
max. power input 3,6 kW
nominal voltage 400/3N V
current 7,8 A
input in stand by mode 5 W
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.