KINEMATIC VISCOCITY WATER BATH 20 °C-100 °C
For Kinematic Viscosity from 20 °C to 100 °C
ASTM D445/D446, ISO 3104/3105
The TR Series constant temperature bath accommodates viscometer tubes requiring a bath depth up to 18" and provides reliable temperature control from 20 °C to 100 °C . The unit accommodates up to 7 simultaneous viscosity determinations.
- Common Applications
- Formulated oil analysis
- Hydraulic oil analysis
- Additive analysis
- Marine fuel testing
- Base stock analysis
- Light and heavy fuel testing
- Waxes/paraffin
- Crude oil testing
- Glycols
2 Models Available
- TR-518 Constant Temperature Bath (120 V)
- TR-518 Constant Temperature Bath (220 V)
Precise and accurate temperature control
- Temperature range: 20 °C to 100 °C ± 0.01 °C*
- Temperature selection is variable between 20 °C and 100 °C or selectable at ten pre-set temperatures (20 °C,25 °C,30° C,
40 °C,50 °C,60 °C,70 °C,80 °C,90 °C,and 100 °C) - Meets ASTM D445 temperature sensitivity requirements for kinematic viscosity determination
- Motor driven stirrer ensures a uniform bath temperature
Safe operation
- Bath thermistor senses over-temperature fault and removes power
- Thermistor disconnection results in removal of bath heater power
- Liquid level sensor prevents operation if bath liquid is not at a safe operating level
Efficient,user-friendly design
- Seven glass viscometer ports (each 2” in diameter) and two thermometer ports (each 0.4” in diameter) allow up to 7 simultaneous viscosity determinations**
- Two electric heating elements quickly heat bath media to the selected temperature
- Stainless steel baffle provides a plain reflective viewing background
- Convenient drawer provides easy access to electrical controls
Reliable performance
- Well tested CANNON bath platform offers reliability and dependable support
- All wetted components are manufactured from high quality stainless steel or glass
- The heavy duty aluminum frame offers a corrosion-resistant epoxy finish
* Test temperatures < 40 °C and high ambient temperature test environments require an optional bath cooler
** Other port configurations are available as a special order
SPECIFICATION:
UNIT WEIGHT | 22 kg (48 lb) |
---|---|
DIMENSIONS (W X D X H) | 40.7 cm × 36.2 cm × 76.2 cm (16 in × 14.25 in × 30 in) |
SHIPPING DIMENSIONS (W X D X H) | 76.2 cm × 58.4 cm × 55.9 cm (30.0 in × 23.0 in × 22.0 in) |
SHIPPING WEIGHT (WITH ALL ITEMS) | 36.3 kg (80 lb) |
SAMPLE CAPACITY | 7 |
TEMPERATURE RANGE AND ACCURACY | 20 °C to 100 °C,± 0.01 °C*,* Test temperatures < 40 °C and high ambient temperature test environments require an optional bath cooler |
OPERATING CONDITIONS | 15 °C to 30 °C,10% to 75% relative humidity (non-condensing),Installation Category II; Pollution Degree 2 |
ELECTRICAL SPECIFICATIONS | 120 Vac,50/60 Hz; 240 Vac,50/60 Hz; 1000 watts power consumption |
COMPLIANCE | CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc,60 sec.); ROHS |
BATH VOLUME | 23.8 L (6.3 gal) |
BATH HEIGHT | 18 in |