Ecocell 22 Eco
ECOCELL - ECO Line
Volumes: 22, 55, 111, 222, 404, 707 litres Working temperature: 5 °C above ambient temperature up to 250 °C (or up to 300°C as option)
The highly cost effective heating oven series for simple drying processes. The ECOCELL-line produces no noise and there is very
soft air convection within the chamber.
* Microprocessor control Fuzzy Logic
* 3-conductors PT-100 sensors
* Integrated RS 232 for Warmcomm or print
* 3 inch transreflective LCD COG display
* Keyboard block of unauthorized access
* Integrated USB Device for Warmcomm
* Electronical security device type 2
* Galvanized sheet metal painted housing
* Inner walls from stainless steel
* Exhaust duct
* 2 chrome plated trays included
* Click out inner walls
* Acoustic and visual alarm
* Up to 2 segments, up to 99 cycles
* 9 programs
* Real time feature
ECOCELL 22 ECO
- Convincingly economical laboratory drying oven
- Chamber volume 22 litres
- Temperature range from 5°C above ambient to 250/300°C
- Natural fine airflow in the work area, noiseless operation
- High-quality and safe drying, heating and tempering of common laboratory materials in the laboratory and industrial areas
- High operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- Output for connecting a printer or PC
- Possibility of validation (IQ, OQ)
Specification
Interior of stainless steel material No. 1.4301
Volume: 22 l
Size D: 350 mm
Size W: 240 mm
Size H: 300 mm
Tray Standard equipment : 2 pcs
Storage area (w × d): 185 × 265 mm
Racks: 4 max. no
Distance between guide rails 60 mm
Maximal weight of the load *)
Inside the oven: 25 kg/Oven
Per tray: 10 kg/tray
Door : 1 pcs
External dimensions (including door and handle)
Depth 560 mm
Diameter of the air branch outer/inner 52/49 mm
Width 406 mm
Package dimensions (three layers carton)
Depth: 720 mm
Width: 500 mm
Depth: 810 mm
Weight bruto: 36 kg
Weigh Nett: 31 kg
Working temperature (beginning of the regulation): from 5 °C above ambient temp. to: 250 °C
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door
Time variation cca (±): 1 °C
Space deviation cca (±) % of the reached temperature: 2,7 %
Time required to reach 250 °C with closed air flap and voltage 230 V: 54 min
Heat emission at 250 °C: 300 W
Air exchange speed at 250 °C: 6 per hour
Electricity - mains 50/60 Hz:
Max. power input: 0,9 kW
Nominal voltage: 230 V
Current: 4 A
Input in stand by mode: 5 W
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate Ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved. The values may differ depending on specific charge and media parameters.